Carbon-related data measurements methods and instrumentation. pH Seawater samples were drawn from the PVC bottles with a 25-cm length of silicon tubing. One end of the tubing was fit over the petcock of the PVC bottle and the other end was attached over the opening of a 10-cm glass spectrophotometric cell. The spectrophotometric cell was rinsed three to four times with a total volume of approximately 200 mL of seawater; the Teflon(tm) endcaps were also rinsed and then used to trap a sample of seawater in the glass cell. While drawing the sample, care was taken to make sure that no air bubbles were trapped within the cell. Seawater pH was measured using a three-wavelength spectrophotometric procedure (Byrne, 1987) and the indicator calibration of Clayton and Byrne (1993). The indicator was a 8.0-mM solution of Kodak(tm) m-cresol purple sodium salt (C21H17O5Na) in a 10% ethanol solution; the absorbance ratio of the concentrated indicator solution (RI = 578A/434A) was 1.00. All absorbance ratio measurements were obtained in the thermostatted (25.0 +/- 0.05°C) cell compartments of HP 8453 diode array spectrophotometers. Periodically the spectrophotometric cells were cleaned with a 1 N HCl solution to preclude biological growth. Measurements of pH were taken at 25.0C on the total hydrogen ion concentration ([H+]t) scale, in mol/kg soln. Total Carbon Dioxide (TCO2) The DIC analytical equipment was set up in a seagoing container modified for use as a laboratory. The analysis was done by coulometry; two analytical systems were used simultaneously on the cruise, each consisting of a coulometer (UIC, Inc.) coupled with a SOMMA (Single Operator Multiparameter Metabolic Analyzer) inlet system developed by Ken Johnson (Johnson et al., 1985,1987,1993; Johnson, 1992) of Brookhaven National Laboratory (BNL). Pipette volume was determined based on the procedures described in Handbook of Methods for CO2 Analysis (DOE, 1994). In the coulometric analysis of DIC, all carbonate species are converted to CO2 (gas) by addition of excess hydrogen to the seawater sample, and the evolved CO2 gas is carried into the titration cell of the coulometer, where it reacts quantitatively with a proprietary reagent based on ethanolamine to generate hydrogen ions. These are subsequently titrated with coulometrically generated OH-. CO2 is thus measured by integrating the total charge required to achieve this. Samples were drawn from the PVC bottles into cleaned, precombusted 500-mL Pyrex(tm) bottles using Tygon(tm) tubing according to procedures outlined in the Handbook of Methods for CO2 Analysis (DOE, 1994). Bottles were rinsed once and filled from the bottom, overflowing half a volume, and care was taken not to entrain any bubbles. The tube was pinched off and withdrawn, creating a 5-mL headspace, and 0.2 mL of saturated HgCl2 solution was added as a preservative. The sample bottles were sealed with glass stoppers lightly covered with Apiezon-L(tm) grease, and were stored at room temperature for a maximum of 12 hours prior to analysis. The coulometers were calibrated by injecting aliquots of pure CO2 (99.995%) by means of an 8-port valve outfitted with two sample loops that had been calibrated at BNL (Wilke, 1993). All DIC values were corrected for dilution by 0.2 mL of HgCl2; total water volume was 540 mL. The correction factor used was 1.00037. The instruments were calibrated at the beginning, middle, and end of each coulometer cell solution with a set of the gas loop injections. CRMs (Batch 29) were provided by Dr. Andrew Dickson (SIO), and was analyzed on both instruments over the duration of the cruise. The CRM certified value was 1902.54 +/-1.05 (n=14). The overall accuracy and precision for the CRMs on both instruments combined was -1.1 +/-0.9 (n=153). Replicate measurements from different PVC bottles tripped at the same depth, along with replicate measurements from the same PVC bottle was within +/-1.9 µmol/kg DIC. DIC data reported for this cruise have been corrected to the Batch 29 CRM value by adding the difference between the certified value and the mean shipboard CRM value (certified value - shipboard analyses) on a per instrument/per leg basis. Total Alkalinity (TA) The titration system used to determine TA consisted of a Metrohm 665 Dosimat(tm) titrator and an Orion(tm) 720A pH meter controlled by a personal computer (Millero et al., 1993). The acid titrant, in a water-jacketed burette, and the seawater sample, in a water-jacketed cell, were kept at 25 +/- 0.1°C with a Neslab(tm) constant-temperature bath. The plexiglass water-jacketed cells were similar to those used by Bradshaw et al. (1988), except that a larger volume (200 mL) was used to increase the precision. The cells had fill and drain valves with zero dead-volume to increase the reproducibility of the cell volume. The HCl solutions used throughout the cruise were made, standardized, and stored in 500-mL glass bottles in the laboratory for use at sea. The 0.2487 M HCl solutions were made from 1 M Mallinckrodt(tm) standard solutions in 0.45 M NaCl to yield an ionic strength equivalent to that of average seawater (0.7 M). The acid was independently standardized using a coulometric technique (Taylor and Smith, 1959; Marinenko and Taylor, 1968) by the University of Miami and by Dr. Dickson. The two standardization techniques agreed to +/-0.0001 N. The volume of HCl delivered to the cell is traditionally assumed to have a small uncertainty (Dickson, 1981) and is equated with the digital output of the titrator. Calibrations of the Dosimat(tm) burettes with Milli Q(tm) water at 25C indicated that the systems deliver 3.000 mL (the value for a titration of seawater) to a precision of 0.0004 mL. This uncertainty resulted in an error of 0.4 µmol/kg in TA. Internal consistency of each cell was checked before, during, and after the cruise by titrating CRM Batches 29 and 30 prepared by Dr. Dickson. The TA of CRM was determined by open cell (weighed) titration in the laboratory prior to the cruise and was found to be 2184.8 +/- 1.3 µmol/kg (n= 15) and 2201.9 +/- 1.0 µmol/kg (n = 21), respectively. A total of 85 CRM measurements made at sea yielded 2173.8 +/- 1.6 µmol/kg for Batch 29 and 2190.8 +/- 1.7 µmol/kg for Batch 30 on three different cells. This offset was due to changes in the volume of the cells. All TA data have been corrected to laboratory CRM values for each cell and each leg.